Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Graphtec WX1200A: (1 year) CAL VER /5500 DATE: 03-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 23 NUMBER OF LINES: 132 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Western Graphtec WX1200A Specifications # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- R N P F V 1.003 ASK+ X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: 20 min. - REQUIRED 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <75% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD INITIAL SETUP 1.008 DISP Load a sheet of graph paper (having dimensions at least 1.008 DISP 20cm x 15cm) and align the paper to be parallel in both 1.008 DISP the X and Y directions. 1.009 HEAD {X AMPLIFIER} 1.010 DISP Connect the 5500A NORMAL HI and LO to the UUT X INPUT. 1.011 DISP Set the X RANGE to .5mV/cm. 1.012 5500 0mV S 2W 1.013 DISP Using the X POSITION control, align the pen to a zero 1.013 DISP reference point. 1.014 HEAD X AMPLIFIER: {.5mV Range} 1.015 MESS Slew the 5500A for exactly 20cm of deflection. 1.015 MESS 1.016 5500 0.5 10.00mV 0.5% 2W 2.001 HEAD X AMPLIFIER: {1mV Range} 2.002 DISP Set the X RANGE to 1mV/cm. 2.003 5500 1 20.00mV 0.5% 2W 3.001 HEAD X AMPLIFIER: {2mV Range} 3.002 DISP Set the X RANGE to 2mV/cm. 3.003 5500 2 40.00mV 0.5% 2W 4.001 HEAD X AMPLIFIER: {5mV Range} 4.002 DISP Set the X RANGE to 5mV/cm. 4.003 5500 5 100.00mV 0.5% 2W 5.001 HEAD X AMPLIFIER: {10mV Range} 5.002 DISP Set the X RANGE to 10mV/cm. 5.003 5500 10 200.00mV 0.5% 2W 6.001 HEAD X AMPLIFIER: {20mV Range} 6.002 DISP Set the X RANGE to 20mV/cm. 6.003 5500 20 400.00mV 0.5% 2W 7.001 HEAD X AMPLIFIER: {50mV Range} 7.002 DISP Set the X RANGE to 50mV/cm. 7.003 5500 50 1000.00mV 0.5% 2W 8.001 HEAD X AMPLIFIER: {.1V Range} 8.002 DISP Set the X RANGE to .1V/cm. 8.003 5500 0.1 2.00V 0.5% 2W 9.001 HEAD X AMPLIFIER: {.2V Range} 9.002 DISP Set the X RANGE to .2V/cm. 9.003 5500 0.2 4.00V 0.5% 2W 10.001 HEAD X AMPLIFIER: {.5V Range} 10.002 DISP Set the X RANGE to .5V/cm. 10.003 5500 0.5 10.00V 0.5% 2W 11.001 HEAD X AMPLIFIER: {5V Range} 11.002 DISP Set the X RANGE to 5V/cm. 11.003 5500 5 100.00V 0.5% 2W 12.001 HEAD {Y AMPLIFIER} 12.002 DISP Set the X RANGE to .5mV/cm. 12.003 DISP Connect the 5500A NORMAL HI and LO to the UUT Y INPUT. 12.004 DISP Set the Y RANGE to .5mV/cm. 12.005 5500 0mV S 2W 12.006 DISP Using the Y POSITION control, align the pen to a zero 12.006 DISP reference point. 12.007 HEAD Y AMPLIFIER: {.5mV Range} 12.008 MESS Slew the 5500A for exactly 15cm of deflection. 12.008 MESS 12.009 5500 0.5 7.50mV 0.5% 2W 13.001 HEAD Y AMPLIFIER: {1mV Range} 13.002 DISP Set the Y RANGE to 1mV/cm. 13.003 5500 1 15.00mV 0.5% 2W 14.001 HEAD Y AMPLIFIER: {2mV Range} 14.002 DISP Set the Y RANGE to 2mV/cm. 14.003 5500 2 30.00mV 0.5% 2W 15.001 HEAD Y AMPLIFIER: {5mV Range} 15.002 DISP Set the Y RANGE to 5mV/cm. 15.003 5500 5 75.00mV 0.5% 2W 16.001 HEAD Y AMPLIFIER: {10mV Range} 16.002 DISP Set the Y RANGE to 10mV/cm. 16.003 5500 10 150.00mV 0.5% 2W 17.001 HEAD Y AMPLIFIER: {20mV Range} 17.002 DISP Set the Y RANGE to 20mV/cm. 17.003 5500 20 300.00mV 0.5% 2W 18.001 HEAD Y AMPLIFIER: {50mV Range} 18.002 DISP Set the Y RANGE to 50mV/cm. 18.003 5500 50 750.00mV 0.5% 2W 19.001 HEAD Y AMPLIFIER: {.1V Range} 19.002 DISP Set the Y RANGE to .1V/cm. 19.003 5500 0.1 1.50V 0.5% 2W 20.001 HEAD Y AMPLIFIER: {.2V Range} 20.002 DISP Set the Y RANGE to .2V/cm. 20.003 5500 0.2 3.00V 0.5% 2W 21.001 HEAD Y AMPLIFIER: {.5V Range} 21.002 DISP Set the Y RANGE to .5V/cm. 21.003 5500 0.5 7.50V 0.5% 2W 22.001 HEAD Y AMPLIFIER: {5V Range} 22.002 DISP Set the Y RANGE to 5V/cm. 22.003 5500 5 75.00V 0.5% 2W 23.001 END